- Easy to install 1/8 DIN plastic casing.
- Clear LCD displays pH, mV or temperature with backlight.
- Universal power supply input from 100-240VAC.
- Automatic buffer recognition of US and NIST buffer sets. Calibrated data is
stored in memory and ready to for use on power up.
- Automatic or manual temperature compensation.
- Displays absolute and relative ORP.
- Two sets of programmable on/off relays for pH or ORP (3621).
- Isolated and reversible 4-20mA current output for pH or ORP allows it to be used as a proportional output (3631).
- RoHS compliant.
Specifications
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Specifications
|
Range
|
Resolution
|
Accuracy
|
|
|
|
|
|
|
|
|
|
|
|
|
pH
|
-2.00~16.00pH
|
0.01pH ±0.1%
|
±1 digit
|
|
|
|
|
mV (absolute / relative)
|
-1999~+1999mV
|
1mV / 1mV
|
±0.1%±1 digit
|
|
|
|
|
Temperature
|
-10.0~120.'C
|
0.1'C
|
±0.3'C
|
|
|
|
|
pH Buffer Recognition
|
US (4.01, 7.00, 10.01) or NIST (4.00, 6.86, 9.18), user selectable.
|
|
|
|
|
pH Electrode Offset Recognition
|
±100mV at pH 7.00 / -91.7mV to +108.3mV at pH 6.86.
|
|
|
|
|
pH Electrode Slope Recognition
|
±30% at pH 4.00, 4.01, 9.18 or 10.01.
|
|
|
|
|
|
Temperature Compensation
|
Manual / Auto -10.0~120.01 (Thermistor, 10K ohm / PT-1000).
|
|
|
|
|
Input Impedance
|
>1013 ohms.
|
|
|
|
|
|
|
Input select
|
User selectable pH or ORP.
|
|
|
|
|
|
Control Type / Relay Output (3621)
|
2 ON/OFF controls /5A at 115V AC or 2.5A at 220V AC, resistive load only.
|
|
|
|
|
Current Output Range / Scale(3631)
|
4 to 20 mA (isolated) / Liner
|
|
|
|
|
|
Maximum Load
|
500 ohm
|
|
|
|
|
|
|
Isolation Voltage
|
500 VDC
|
|
|
|
|
|
|
Display
|
LCD, with backlighting.
|
|
|
|
|
|
|
Power
|
100~240VAC at 50/60 HZ.
|
|
|
|
|
|
Ambient Temperature
|
Range From 0~501.
|
|
|
|
|
|
|
Case
|
1/8 DIN case, 90mm depth.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Accessories
|
|
|
|
|
|
|
|
IP-600-10
|
For wastewater (pH, glass body, PPS housing, NPT3/4’’ thread, 20 m cable).
|
|
|
|
|
IP-600-9 TH
|
For wastewater (pH, glass body, PPS housing, NPT3/4’’ thread, 5 m cable, with 10K thermistor).
|
|
|
|
|
IR-500-8
|
For wastewater (ORP, glass body, PPS housing, NPT3/4’’ thread, 5 m cable).
|
|
|
|
|
009A
|
Protective submersion housing.
|
|
|
|
|
|
012-150-00
|
Waterproof enclosure
|
|
|
|
|
|
|
|
|
|
|
|
|
|